Product Specifications for Series 1399: DC to 100 MHzVoltage and Frequency Ranges
Use recommended voltage ranges to retain optimum accuracy. Calibration Uncertainty
* Expressed in ppm unless otherwise indicated Measurements Reference 23 °C ± 1 °C. All 1399 series converters are furnished complete with a Ballantine production calibration certificate directly traceable to NIST. INPUT IMPEDANCE: 200 W ± 10% per nominal Volt. Input Capacitance: 10 pF nominal. Thermoelement: Nominal Heater Current: 5 mA rms, Max. Heater Current: 7.5 mA rms. Couple Output: 9 mV ± 12% at nominal heater current. Couple Resistance: 30 W nominal Maximum DC Reversal Error: ± 0.005% (3.5 to 5 mA heater current) ± 0.01% (2.5 to 3.5 mA heater current) Couple Output Voltage to Heater (ground): ± 50 V pk max. Connectors: Signal Input: Type N Female Environmental Size & Weight Revised: April 12, 2006. |
|