Model 16201 5/10 Turn Current coil

Introduction:
The 5/10 turn 16201 Current coil is used as a tool for
calibrating clamp type current meters that operate by two different principles - as
current transformers (AC only), and by the Hall Effect (both AC and DC). As it is
typically impractical to calibrate 1000A rated current clamp meters with a 1000A source,
one can easily use 10 turns of the 16201 coil in conjunction with a 1620A current source calibrator to effectively
multiply the current of the current source calibrator by a factor of 10. Now it is
easy to support the calibration and verification of these clamp-on type current
meters.
Using the Coil:
Clamp-on current meters operate as current transformers, with varying
degrees of magnetic coupling between the primary and secondary. The position of the
clamp meter with respect to the cable also affects the magnetic coupling between the
primary and the secondary of the current transformer, which causes variation in reading of
the current meter. This is important to understand in order to make the most
accurate and repeatable measurements.
Product Comparison
Features
|
Ballantine 16201
|
Fluke 5500A
|
| Coils |
5 and 10 Turns of 5 gauge wire |
50 Turns of 16 gauge wire |
| Maximum Input Current |
60 A rms continuous
100A rms, 60 minutes |
11 A rms continuous
20 A rms, 2 minutes |
| Maximum Duty Cycle Derating |
< 60A, continuous
100A, 60 min. ON, 30 min. Off |
<11 A, continuous
>11 A, 2 minutes ON, 8 minutes OFF |
| Maximum Voltage |
4 V DC, 3.5 Vrms AC |
3 V rms |
| Frequency of Operation |
DC, 10- 1000 Hz, (should not exceed rms
Voltage rating).
For the 16201 Current Output;
DC, 10-800 Hz, 0-100A
This specification assumes that a clamp meter is attached. Some clamp meters
may limit the 16201 Current Output drive capability to less than that above |
DC, 45-440 Hz (should not exceed rms
Voltage rating).
For the 5500A Current Output;
DC, 45-65 Hz 0-11A
65-440 Hz, 0-4 A
This specification assumes that a Fluke 801-600 Clamp Meter or equivalent is attached.
Other clamp meters may limit the 5500A Current Output drive capability to less than
that above. |
| Uncertainty due to Clamp Meter/Coil interaction |
± 0.25% of effective output +0.05A for
toroidal wound current clamps, such as the Fluke 801, 801-1000, Lem Hemme LH240, LH1040,
LH2015 and LH2040. ± 0.50% of effective output +0.5A for current clamps
like the Fluke 801-KV, 801-410, 801-500, 801-1010, Fluke 31 and Fluke 33 or equivalent. |
± 0.25% of effective output +0.05A for
toroidal wound current clamps, such as the Fluke 801 and 801-1000. ±
0.50% of effective output +0.5A for current clamps like the Fluke 801-KV, 801-410,
801-500, 801-1010, Fluke 31 and Fluke 33 or equivalent. |
| Minimum ID of Clamp Jaws |
3.8 cm (1.5in.) x 2 cm (.8 in.) |
2.54 cm (1 in.) |
| |
1 year absolute
uncertainty, tcal ± 5° C, ± (% of effective current +A) |
Range
|
Torodial-Wound
Current Clamps |
Other Current
Clamps |
| DC, 0-100A |
0.25% |
0.05A |
0.50% |
0.50A |
| 45-65 Hz, 1-100A |
0.25% |
0.05A |
0.50% |
0.50A |
| 65-880 Hz, 100A |
0.27% |
0.1A |
0.51% |
0.51A |
|