Model 16201 5/10 Turn Current coil

16201

10 Turn Current coil Domestic Price--$1095.00 [Order] [Checkout]

Introduction:

The 5/10 turn 16201 Current coil is used as a tool for calibrating clamp type current meters that operate by two different principles - as current transformers (AC only), and by the Hall Effect (both AC and DC).  As it is typically impractical to calibrate 1000A rated current clamp meters with a 1000A source, one can easily use 10 turns of the 16201 coil in conjunction with a 1620A current source calibrator to effectively multiply the current of the current source calibrator by a factor of 10.  Now it is easy to support the calibration and   verification of these clamp-on type current meters.   

Using the Coil:

Clamp-on current meters operate as current transformers, with varying degrees of magnetic coupling between the primary and secondary.  The position of the clamp meter with respect to the cable also affects the magnetic coupling between the primary and the secondary of the current transformer, which causes variation in reading of the current meter.  This is important to understand in order to make the most accurate and repeatable measurements.

Product Comparison

Features
Ballantine 16201
Fluke 5500A
Coils 5 and 10 Turns of 5 gauge wire 50 Turns of 16 gauge wire
Maximum Input Current 60 A rms continuous
100A rms, 60 minutes
11 A rms continuous
20 A rms, 2 minutes
Maximum Duty Cycle Derating < 60A, continuous
100A, 60 min. ON, 30 min. Off
<11 A, continuous
>11 A, 2 minutes ON, 8 minutes OFF
Maximum Voltage 4 V DC, 3.5 Vrms AC 3 V rms
Frequency of Operation DC, 10- 1000 Hz, (should not exceed rms Voltage rating).
For the 16201 Current Output;
DC, 10-800 Hz, 0-100A

This specification assumes that a clamp meter is attached.  Some clamp meters may limit the 16201 Current Output drive capability to less than that above
DC, 45-440 Hz (should not exceed rms Voltage rating).
For the 5500A Current Output;
DC, 45-65 Hz 0-11A
65-440 Hz, 0-4 A
This specification assumes that a Fluke 801-600 Clamp Meter or equivalent is attached.   Other clamp meters may limit the 5500A Current Output drive capability to less than that above.
Uncertainty due to Clamp Meter/Coil interaction ± 0.25% of effective output +0.05A for toroidal wound current clamps, such as the Fluke 801, 801-1000, Lem Hemme LH240, LH1040, LH2015 and LH2040.

± 0.50% of effective output +0.5A for current clamps like the Fluke 801-KV, 801-410, 801-500, 801-1010, Fluke 31 and Fluke 33 or equivalent.

± 0.25% of effective output +0.05A for toroidal wound current clamps, such as the Fluke 801 and 801-1000.

± 0.50% of effective output +0.5A for current clamps like the Fluke 801-KV, 801-410, 801-500, 801-1010, Fluke 31 and Fluke 33 or equivalent.

Minimum ID of Clamp Jaws 3.8 cm (1.5in.) x 2 cm (.8 in.) 2.54 cm (1 in.)
 

1 year absolute uncertainty, tcal ± 5° C, ± (% of effective current +A)

Range

Torodial-Wound Current Clamps

Other Current Clamps

DC, 0-100A 0.25% 0.05A 0.50% 0.50A
45-65 Hz, 1-100A 0.25% 0.05A 0.50% 0.50A
65-880 Hz, 100A 0.27% 0.1A 0.51% 0.51A

 

All prices subject to change without notice.  All prices listed are for domestic sales only, please call our offices to confirm price.