Model 6127B Programmable Oscilloscope Calibrator pdficon(1).gif (224 bytes)


Product Summary6127B.jpg (20181 bytes)

The BLI 6127B is an accurate and simple way to calibrate a wide variety of oscilloscopes from low frequency units to instruments bandwidths over 1 GHz. The 6127B provides a variety of signals of precise voltage, current, timing and waveshape. The 6127B can be operated manually or automatically using the IEEE-488 interface and Ballantines automated software SCOPE-CAL™. Time saving controls include computation of percent deviation, repeat step increment, self checking and warning of invalid entries.

Ballantine has over 300 procedures for many popular scopes on the market. For information on our Procedure Library click here


Product Features

bulletMaximum adaptability. Checks any scope from general purpose to advanced 1 GHz units.
bulletCuts calibration time and cost. Halves test time. Speeds throughput to 5 times when programmed.
bulletEasy conversion to SCOPE-CALTM, Ballantine's Automated Oscilloscope Calibration Software
bulletOver 300 free procedures
bulletManual or IEEE 488 GPIB operation.
bulletUser friendly µP based panel computes calibration errors, self checks, flags invalid entries.


Product Specifications

Volts/Division Mode

Used for calibration of vertical display accuracy and attenuator compensation.

High Impedance Output (into 1 MW): ): ): ): Range: 40 µV to 200 V in 1, 2, 5 steps with multiplier
Low Impedance Output (into 50
Low Impedance Output (into 50
Low Impedance Output (into 50
Low Impedance Output (into 50 W): ): Range: 40 µV to 5 V in 1, 2, 5 steps with multiplier
Multiplier: 1, 2, 3, 4, 5, 6, 8, 10 divisions
Amplitude Accuracy: ±0.25% of reading ±1 µV
Variable Amplitude Deviation: Variable Range of ±9.9%. Digital deviation meter provides resolution of 0.1%.
Waveforms: Selectable outputs for all output levels. Square wave of 10 Hz, 100 Hz, 1 kHz, and 10 kHz. DC positive polarity. DC negative polarity.

Time/Division Mode

Real Time Markers for calibration of oscilloscope timebase accuracy and checking high frequency triggering.
Output Marker: 500 psec to 5 sec in 1, 2, 5 sequence
Marker Amplitude: ³1 V pk into 50 W, 2 nsec to 5 sec >350 mV pk-to-pk at 1 nsec (1 GHz)
>100 mV pk-to-pk at 500 psec (2 GHz)
Accuracy: ±0.005% of reading ±10 psec (2 nsec to 5 sec), ±0.005% of reading ±20 psec (0.5 nsec to 1 sec)
Variable Time Deviation: Variable deviation range of ±9.9% for marker and trigger period. Digital deviation meter provides resolution of 0.1%.

Timing Reference

Internal Time Base: An internal temperature compensated 10 MHz crystal oscillator provides accuracy of ±0.002% and long term stability to better than 1 part in 10-5 per month after 72 hours and temperature stability of better than 1 part in 10-5 from +4°C to +40°C after a one hour warm-up.
External Timing Reference: Input: 10 MHz, Input Amplitude: 1 V to 10 V rms., Source Impedance: 50 W

Milliamperes/Division (Current Amplitude Mode)

Used to test current probe accuracy through front panel current loop.
Current Ranges: 1 mA to 100 mA in 1, 2, 5 steps with multiplier
Multiplier: 1, 2, 3, 4, 5, 6, 8, 10 divisions
Accuracy: ±0.25% of reading ±2 µA
Frequency: ±DC: Squarewave of 10 Hz, 100 Hz, 1 kHz, and 10 kHz

Calibrator Mode

For use with 61272C Amplitude Comparator Head to test accuracy of oscilloscope's internal calibrator signal.
Amplitude Range: High Impedance: ±40 µV to ±200 V in 1, 2, 5 sequence
Low Impedance (50 Low Impedance (50 W): ): Low Impedance (50 Low Impedance (50
W): ): ): ):
±40 µV to ±5 V in 1, 2, 5 sequence
Accuracy (Amplitude): ±0.25% of reading ±1µV
Accuracy of 50 Accuracy of 50 W Impedance: Impedance: ±1%
Frequency of UUT Calibrator Waveform: DC to >2 MHz
Frequency of Calibration Waveform: 1 kHz squarewave
Comparison Mode: Compares output of Volts/Div generator and calibrator of scope under test. Comparison is switched automatically at 100 Hz or alternate on command.

Low Distortion Pulse Mode

Used to test oscilloscope attenuator compensation and input amplifier response.

Low Edge (Amplitude Mode) High Edge (Amplitude Mode)
Range 20 mV to 1 V pk-to-pk in 1, 2, 5 steps 1.2 V to 100 V pk-to-pk in 1, 2, 5 steps
Multiplier 1, 2, 3, 4, 5, 6, 8, 10 divisions 1, 2, 3, 4, 5, 6, 8, 10 divisions
Risetime £1.3 nsec from 100 mV to 1 V into 50 W

£1.6 nsec from 20 mV to 50 mV into 50 W

<100 nsec
Load Resistance n/a £1 MW load shunted by less than 18 pF
Polarity Positive transition from negative voltage to ground Positive transition from negative voltage to ground
Output Impedance 50 W ±1% n/a
Waveshape Squarewave with 50% duty cycle Squarewave with 50% duty cycle. Leading edge aberrations within 2% of pk-to-pk.
Aberrations ±2% of squarewave amplitude after first 1.5 nsec ±2% of squarewave amplitude after first 1.5 nsec
Long Term Flatness Droop and tilt within ±0.5% Droop and tilt within ±0.5% after first 500 nsec
Frequency 10 Hz to 1 MHz to 6 decade steps 10 Hz to 100 kHz in 5 decade steps

Fast Rise Pulse

The Fast Rise Pulse Head provides a fast rise, low distortion pulse used for testing high bandwidth vertical amplifiers.
Amplitude: 1.1 V pk-to-pk, ±5%, adjustable ±10% into 50 W
Risetime: <180 psec into 50 W
Waveshape: Output is positive squarewave 50% duty cycle. Leading edge aberrations are less than ±2% of amplitude not exceeding 4% pk-to-pk.
Frequency: 10 Hz to 1 MHz in 6 decade steps

Trigger Output

Slaved to frequency of selected mode and marker. Time/Division Mode, trigger rate fixed at 100 nsec for marker outputs of 100 nsec and faster.
Modes: OFF, x1, +10, +100
Amplitude: ³1 V into 50 W

Bus Address

Selected by switch accessible on rear panel and displayable on front panel.

Power

100/120/220/240 V ac, ±10%, 50 to 400 Hz, single phase sinusoidal, ~80 W.

Environmental

Meets or exceeds MIL-T-28800, Type III, Class 5 requirements, including Reliability and Maintainability. Satisfies EMI requirements per MIL-STD-461. CE-03, CS01, CS06, RE02, RE03.
Temperature: Operating: 0°C to 50°C, Storage: -55°C to 75°C
Humidity: Operating: 0 to 95% RH to 40°C, Storage: 0 to 95% RH to 50°C
Altitude: Operating: 3 km (10,000 ft), Storage: 15 km (50,000 ft)
Vibration: Operating: 0.015 inches pk-to-pk, 10 Hz to 55 Hz
Shock: Non-Operating: 30 G, 1/2 sine

Size & Weight

133mm (5.22 in) H x 425mm (16.75 in) W x 484mm (19.1 in) D
Weight: 15.2 kg (33.5 lbs.); Shipping: 17.7 kg (39 lbs.)

 


Product Ordering Information

Model Description
6127B Programmable Oscilloscope Calibrator with IEEE-488 and 61278 (Requires additional selection of appropriate Head)
90-10310-5B 6127B Manual, Volume 1: Domestic Price - $75.00 [Order] [Checkout]
Heads for 500 MHz Scopes
61252A Fast Rise Pulser
61278 Low Frequency Head
For 1 GHz Scopes
61271C Fast Rise Head
61274A Direct Output Cable
Accessories  
10650A Capacitance Standardizer for Oscilloscopes
61272C Amplitude Comparator Head
61277A Kit Containing 12619A, 12621A (2), 12630A (2), 12680A (2), 12681A, 12682A, 12683A, 12684A, 12685A, 12686A, 12687A, 12688A, 12689A, 12690A, and 89-11103-1A
89-10753-1A Rack Slide Kit
89-11175-1A Protective Front Cover. Storage for two Heads and 61274A
89-11176-1A Protective Rear Cover. Storage for power cord




     