Model 6127B Programmable Oscilloscope Calibrator


Product Summary
The BLI 6127B is an accurate and simple way to calibrate a wide variety of
oscilloscopes from low frequency units to instruments bandwidths over 1 GHz. The 6127B
provides a variety of signals of precise voltage, current, timing and waveshape. The 6127B
can be operated manually or automatically using the IEEE-488 interface and Ballantines
automated software SCOPE-CAL. Time saving
controls
include computation of percent deviation, repeat step increment, self checking and warning
of invalid entries.
Ballantine has over 300 procedures for many popular scopes on the market. For
information on our Procedure Library click here



Product Features
 | Maximum adaptability. Checks any scope from general purpose to advanced 1 GHz
units. |
 | Cuts calibration time and cost. Halves test time. Speeds throughput to 5 times
when programmed. |
 | Easy conversion to SCOPE-CALTM, Ballantine's
Automated Oscilloscope Calibration Software |
 | Over 300 free procedures |
 | Manual or IEEE 488 GPIB operation. |
 | User friendly µP based panel computes calibration errors, self checks, flags
invalid entries. |


Product Specifications
Volts/Division Mode
Used for calibration of vertical display
accuracy and attenuator compensation.
High Impedance Output (into 1 MW): ): ): ): Range: 40 µV to 200 V in 1, 2, 5
steps with multiplier
Low Impedance Output (into 50
Low Impedance Output (into 50
Low Impedance Output (into 50
Low Impedance Output (into 50 W):
): Range: 40 µV to 5 V in 1, 2, 5 steps with multiplier
Multiplier: 1, 2, 3, 4, 5, 6, 8, 10 divisions
Amplitude Accuracy: ±0.25% of reading ±1 µV
Variable Amplitude Deviation: Variable Range of ±9.9%. Digital deviation meter
provides resolution of 0.1%.
Waveforms: Selectable outputs for all output levels. Square wave of 10 Hz, 100 Hz,
1 kHz, and 10 kHz. DC positive polarity. DC negative polarity.
Time/Division Mode
Real Time Markers for calibration of oscilloscope timebase
accuracy and checking high frequency triggering.
Output Marker: 500 psec to 5 sec in 1, 2, 5 sequence
Marker Amplitude: ³1 V pk into 50
W, 2 nsec to 5 sec >350 mV pk-to-pk at 1 nsec (1 GHz)
>100 mV pk-to-pk at 500 psec (2 GHz)
Accuracy: ±0.005% of reading ±10 psec (2 nsec to 5 sec), ±0.005% of reading ±20
psec (0.5 nsec to 1 sec)
Variable Time Deviation: Variable deviation range of ±9.9% for marker and trigger
period. Digital deviation meter provides resolution of 0.1%.
Timing Reference
Internal Time Base: An internal temperature
compensated 10 MHz crystal oscillator provides accuracy of ±0.002% and long term
stability to better than 1 part in 10-5 per month after 72 hours and
temperature stability of better than 1 part in 10-5 from +4°C to +40°C after
a one hour warm-up.
External Timing Reference: Input: 10 MHz, Input Amplitude: 1 V to 10 V rms., Source
Impedance: 50 W
Milliamperes/Division (Current Amplitude Mode)
Used to test current probe accuracy through front panel
current loop.
Current Ranges: 1 mA to 100 mA in 1, 2, 5 steps with multiplier
Multiplier: 1, 2, 3, 4, 5, 6, 8, 10 divisions
Accuracy: ±0.25% of reading ±2 µA
Frequency: ±DC: Squarewave of 10 Hz, 100 Hz, 1 kHz, and 10 kHz
Calibrator Mode
For use with 61272C Amplitude Comparator Head to test
accuracy of oscilloscope's internal calibrator signal.
Amplitude Range: High Impedance: ±40 µV to ±200 V in 1, 2, 5 sequence
Low Impedance (50 Low Impedance (50 W): ):
Low Impedance (50 Low Impedance (50 W):
): ): ): ±40 µV to ±5 V in 1, 2, 5 sequence
Accuracy (Amplitude): ±0.25% of reading ±1µV
Accuracy of 50 Accuracy of 50 W
Impedance:
Impedance:
±1%
Frequency of UUT Calibrator Waveform: DC to >2 MHz
Frequency of Calibration Waveform: 1 kHz squarewave
Comparison Mode: Compares output of Volts/Div generator and calibrator of scope
under test. Comparison is switched automatically at 100 Hz or alternate on command.
Low Distortion Pulse Mode
Used to test oscilloscope attenuator compensation and input
amplifier response.
|
Low Edge (Amplitude Mode) |
High Edge (Amplitude Mode) |
| Range |
20 mV to 1 V pk-to-pk in 1, 2, 5
steps |
1.2 V to 100 V pk-to-pk in 1, 2, 5
steps |
| Multiplier |
1, 2, 3, 4, 5, 6, 8, 10 divisions |
1, 2, 3, 4, 5, 6, 8, 10 divisions |
| Risetime |
£1.3 nsec from 100 mV to 1 V into 50 W £1.6 nsec from 20 mV to 50 mV
into 50 W |
<100 nsec |
| Load Resistance |
n/a |
£1 MW
load shunted by less than 18 pF |
| Polarity |
Positive transition from negative
voltage to ground |
Positive transition from negative
voltage to ground |
| Output Impedance |
50 W ±1% |
n/a |
| Waveshape |
Squarewave with 50% duty cycle |
Squarewave with 50% duty cycle.
Leading edge aberrations within 2% of pk-to-pk. |
| Aberrations |
±2% of squarewave amplitude after
first 1.5 nsec |
±2% of squarewave amplitude after
first 1.5 nsec |
| Long Term Flatness |
Droop and tilt within ±0.5% |
Droop and tilt within ±0.5% after
first 500 nsec |
| Frequency |
10 Hz to 1 MHz to 6 decade steps |
10 Hz to 100 kHz in 5 decade steps |
Fast Rise Pulse
The Fast Rise Pulse Head provides a fast rise, low
distortion pulse used for testing high bandwidth vertical amplifiers.
Amplitude: 1.1 V pk-to-pk, ±5%, adjustable ±10% into 50 W
Risetime: <180 psec into 50 W
Waveshape: Output is positive squarewave 50%
duty cycle. Leading edge aberrations are less than ±2% of amplitude not exceeding 4%
pk-to-pk.
Frequency: 10 Hz to 1 MHz in 6 decade steps
Trigger Output
Slaved to frequency of selected mode and marker.
Time/Division Mode, trigger rate fixed at 100 nsec for marker outputs of 100 nsec and
faster.
Modes: OFF, x1, +10, +100
Amplitude: ³1
V into 50 W
Bus Address
Selected by switch accessible on rear panel and displayable
on front panel.
Power
100/120/220/240 V ac, ±10%, 50 to 400 Hz, single phase
sinusoidal, ~80 W.
Environmental
Meets or exceeds MIL-T-28800, Type III, Class 5
requirements, including Reliability and Maintainability. Satisfies EMI requirements per
MIL-STD-461. CE-03, CS01, CS06, RE02, RE03.
Temperature: Operating: 0°C to 50°C, Storage: -55°C to 75°C
Humidity: Operating: 0 to 95% RH to 40°C, Storage: 0 to 95% RH to 50°C
Altitude: Operating: 3 km (10,000 ft), Storage: 15 km (50,000 ft)
Vibration: Operating: 0.015 inches pk-to-pk, 10 Hz to 55 Hz
Shock: Non-Operating: 30 G, 1/2 sine
Size & Weight
133mm (5.22 in) H x 425mm (16.75 in) W x 484mm (19.1 in) D
Weight: 15.2 kg (33.5 lbs.); Shipping: 17.7 kg (39 lbs.)


Product Ordering Information
| Model |
Description |
| 6127B |
Programmable Oscilloscope
Calibrator with IEEE-488 and 61278 (Requires additional selection of appropriate Head) |
| 90-10310-5B |
6127B Manual, Volume 1:
Domestic Price - $75.00 [Order]
[Checkout] |
| Heads |
for 500 MHz Scopes |
| 61252A |
Fast Rise Pulser |
| 61278 |
Low Frequency Head |
|
For 1 GHz Scopes |
| 61271C |
Fast Rise Head |
| 61274A |
Direct Output Cable |
| Accessories |
|
| 10650A |
Capacitance Standardizer for
Oscilloscopes |
| 61272C |
Amplitude Comparator Head |
| 61277A |
Kit Containing 12619A, 12621A (2),
12630A (2), 12680A (2), 12681A, 12682A, 12683A, 12684A, 12685A, 12686A, 12687A, 12688A,
12689A, 12690A, and 89-11103-1A |
| 89-10753-1A |
Rack Slide Kit |
| 89-11175-1A |
Protective Front Cover. Storage
for two Heads and 61274A |
| 89-11176-1A |
Protective Rear Cover. Storage for
power cord |
|