Model 1605B Thermal Transfer Standard

Product Summary
The BLI 1605B AC/DC Thermal Transfer Standard
is a highly accurate true-rms multi-ranging ac measuring instrument. It serves as a
laboratory standard against which to compare all precision ac instruments. It can provide
absolute transfer measurements to within 20 ppm to 20 kHz, and is the only transfer
standard with a frequency range to 100MHz.


Product Features
 | Precision, true RMS measurement with accuracy traceable to
NIST of typically less than 20 ppm |
 | High Repeatability |
 | Accurate Measurements from 10 Hz to 100 MHz (usable to 250
MHz) |
 | Computer control for automatic testing and report generation |
 | Shorter test time through simpler operation |


Product Specifications
Input:
Frequency Range: ±DC, 10 Hz to 100 MHz; usable from under 5 Hz to over 250 MHz.
Voltage Ranges: ± DC and AC RMS (overrange +10% , underrange -10%; 225
mV min. ,1100 V max.)
Crest Factor: 100:1 of full range; ±450 V pk max. on Low Voltage Input,
± 1500 V pk. max. on High Voltage input.
Input Overload: Fully protected at all frequencies to 10 x range to 1100
V ±(dc + ac rms sinusoidal) applied continuously; overload puts unit in standby mode,
which disconnects input signal from thermo-element until overload is removed.
Transfer Uncertainties:
| NIST traceable 1 year absolute
uncertainties expressed in ppm unless otherwise noted. |
| Voltage |
DC |
100 Hz |
20-20 kHz |
50 kHz |
100 kHz |
1 MHz |
10 MHz |
30 MHz |
100 MHz |
| .5V |
40 |
100 |
45 |
50 |
70 |
220 |
.180% |
.40% |
2.00% |
| 1V |
23 |
100 |
30 |
40 |
70 |
220 |
.180% |
.40% |
1.50% |
| 2V |
16 |
100 |
23 |
35 |
70 |
220 |
.180% |
.40% |
1.50% |
| 4V-32V |
15 |
100 |
23 |
35 |
70 |
220 |
.180% |
.40% |
2.00% |
| 32V-125V |
15 |
100 |
30 |
40 |
70 |
250 |
- |
- |
- |
| 250V |
16 |
- |
30 |
45 |
100 |
- |
- |
- |
- |
| 500V |
16 |
- |
40 |
70 |
120 |
- |
- |
- |
- |
| 1000V |
16 |
- |
50 |
80 |
150 |
- |
- |
- |
- |
Input Impedance: 200 Ohms/Volt.
Input Isolation: Connectors may be floated ±25 V ref. to case ground.
Operating Modes:
Manual Balance: Operator manually null balances the transfer standard in either the normal
AC input to DC output mode or the inverse DC to AC transfer to permit setting input AC
signal to a precise DC level.
Automatic Balance: AC to DC transfer performed only after input signal
rms amplitude drift is verified as less than the following:
AUTO HI ACCURACY mode - 10 ppm within 40 s
AUTO NORM mode - 20 ppm within 20 s
AUTO FAST mode - 40 ppm within 15 s
Manual Transfer: Balance is automatic in the AC input mode, but operator
initiates transfer to DC output based on null meter reading.
Auto-Recycle: Repeats transfer measurement cycle approximately 15 s after
each output READ command: Auto-Recycle is an auto-zero mode, which avoids errors due to
thermal drift.
DC Output: Internally generated DC +0.225 to 1100 V (10 Megohms load) and
averaged by external voltmeter for 10 seconds following the READ command; ripple and noise
95 dB below DC output amplitude or 100 mV, which ever is greater; maximum output current 2
mA.
READ Command: Sent over IEEE-488 bus to trigger external voltmeter to
read transferred DC output after transfer standard DC output has settled; front panel READ
light indicates command.
Interface Bus: IEEE-488 1978.
Functions - SH1, AH1, T10, LEO, T5, L4, SR1, PP0, RL1, DC1, DT0, C0
Bus Controls: Range (12), Balance Mode (5), Operate, Standby Recycle (3), Measure Input,
Measure Output.Isolation: Opto-isolated from measuring circuits; bus digital ground
connected to case.
Environmental: Complies with MIL-T-28800,
Class 5 for shock, vibration, transit drop, humidity to 90%, R.H. non-condensing;
temperature 0 to +50° C operating(+ 10°C to +40°C full specifications) and -40°C to +
75°C storage; altitude 3000 m operating with storage to 15,000 m; fungus resistant; EMI
protected.
Power: 100, 120, 220 and 240 V ac, +8% to-10%; 50 to 400 Hz; 80 VA
Dimensions: Basic unit - 178 mm (7") H x 456 mm (16 3/4") W x 508 mm (20")
D
Enclosure - 203 mm (8") H x 521 mm (20 1/2") W x 527 mm (20 3/4") D
Panel width - 483 mm (19")
Depth behind panel - 470 mm (18 1/2")
Weight: 25 kg (56 lbs.); shipping weight 31.7 kg (70 lbs.)


Product Ordering Information
| Model |
Description |
| 1605B |
Programmable, 225mV to 1100V; ±
DC, 10 Hz to 100 MHz with IEEE-488 and 16056A |
| Accessories |
|
| 16054A |
Maintenance Kit |
| 16056A |
5-Port Precision Tee, Type N-Connectors |
| 16056A-GR |
5-Port Precision Tee, Type GR874 Connectors |
| 16058A |
DP DT DC/DC Transfer Switch |
| 69-10030-0 |
Rack Slides |
|